What does intermediate precision mean in a vape nicotine lab report?
Compare within-run and across-day precision columns using the documented nicotine aerosol method design.
No Vape · Updated

Find the change between the two columns
Read the validation design before trying to interpret a precision percentage. If several samples are prepared and analyzed in a single run, the scatter mainly describes repetitions under that run’s conditions. When validation includes several days of work at the same lab, the precision summary can reflect variation associated with days as well.
Look for the factors actually named. A paper may include changes in analysts or equipment in its validation design. A report that documents different days but says nothing about operators supports a narrower claim about the changes tested.
Read the example in an actual method table
In a published method for nicotine in e-cigarette aerosols, the researchers describe repeating aerosol generation and analysis for validation samples over three consecutive days. Their Table 1 gives separate columns for repeatability and intermediate precision, with relative standard deviation as the unit for both. The same unit makes the numbers look similar while the study design gives the columns different meanings.
The table also reports bias and a tolerance interval. Keep these in their own columns. The validation concerned the chosen method and test conditions. It gives no cross-laboratory comparison, consumer device ranking or personal nicotine measurement.
Read a fictional result without inventing a cause
Imagine a lab checks the same kind of reference preparation on Monday, Tuesday and Wednesday. Results cluster tightly within each day, but the center shifts a little on Wednesday. A within-day summary can look tighter than a calculation that includes all three days. Neither summary alone identifies what caused the shift.
If a caption merely says the result was repeated, ask whether it covers reruns of the same extract or renewed preparation and collection. Repeating only the last instrument step covers less of the process. The answer changes the scope of the precision claim even when the units stay the same.
Keep the scope visible when quoting the report
Record the quantity, repeated step, days or other conditions varied, and what the report leaves open. If you cannot locate the design, use the paper’s own label with that uncertainty rather than supplying an unreported validation detail.
This reading exercise does not choose a performance threshold or offer a correction to any reported concentration. It helps separate evidence that a method held up across particular days from questions about what it measured and whether a different lab would obtain similar results.
Sources and scope
Sources inform the guidance. The worksheets and examples are No Vape editorial suggestions, not tested treatments.